USP, EP and JP TOC requirements compared

USP, EP and JP TOC requirements compared

USP, EP and JP TOC requirements compared

Amscorp Scientific

How USP <643>, Ph. Eur. 2.2.44 and JP 2.59 differ on TOC limits, standards and system suitability, including the 2026 EP changes for sterilised water for injections.

How USP <643>, Ph. Eur. 2.2.44 and JP 2.59 differ on TOC limits, standards and system suitability, including the 2026 EP changes for sterilised water for injections.

USP, EP and JP TOC requirements compared

If your site supplies medicines to more than one market, your pharmaceutical water may have to meet more than one pharmacopoeia. For total organic carbon (TOC), the three that come up most often are the United States Pharmacopeia (USP <643>), the European Pharmacopoeia (Ph. Eur. 2.2.44) and the Japanese Pharmacopoeia (JP 2.59).

For many years, USP and EP were close enough that one set of system suitability standards covered both. That changed on 1 July 2026, when the revised Ph. Eur. texts came into force. JP has always taken its own approach. This article sets out the main differences in plain language, and what they mean for QC and engineering teams in Australia and New Zealand.

If you are still deciding how to measure TOC, our guide to laboratory vs online TOC analysis covers that choice. For the basics, start with what TOC tells you about pharmaceutical water.

The short answer

  • Limits for bulk water line up. USP, EP and JP all work to 0.50 mg/L (500 ppb) of carbon for bulk Purified Water and Water for Injection.

  • USP and EP use the same chemistry, but different reference materials. Both test with sucrose and 1,4-benzoquinone and use the same response efficiency calculation. Since 1 July 2026, EP requires these to be made from EP Chemical Reference Substances (CRS), while USP standards are made from USP Reference Standards.

  • Packaged sterile water now needs TOC under EP too. For sterilised water for injections, Ph. Eur. replaced the old oxidisable substances test with a TOC test, bringing it closer to USP.

  • JP is different again. JP calibrates the analyser with potassium hydrogen phthalate (KHP) and checks oxidation performance with sodium dodecylbenzenesulfonate (SDBS).

Comparison table


USP <643>

Ph. Eur. 2.2.44 (Issue 12.3, from 1 July 2026)

JP 2.59

Scope

Bulk water (Purified Water, Water for Injection) and sterile packaged waters

Method A: bulk Purified Water and Water for Injection. Method B (new): Sterilised Water for Injections in its container

Test for organic carbon in water, used by the JP Purified Water and Water for Injection monographs

Bulk water limit

0.50 mg/L carbon (500 µg C/L target limit response)

0.50 mg/L (500 ppb) carbon, Method A

Not more than 0.50 mg/L for Purified Water and Water for Injection

Packaged or sterile water

Limits depend on nominal container volume. Limit 1: 32 mg/L (5 mL or less), 24 mg/L (over 5 mL to 100 mL), 8 mg/L (over 100 mL), with a higher Limit 2 for each size

Method B uses the same three container sizes. Limit 1: 32, 24 and 8 mg/L. Limit 2: 48, 36 and 12 mg/L. It replaces the old oxidisable substances test

The JP packaged water monographs (for example Sterile Water for Injection in Containers) use a potassium permanganate-reducing substances test rather than a TOC limit

Calibration standard

Not specified

Not specified

Potassium hydrogen phthalate (KHP)

System suitability standards

Sucrose (standard solution) and 1,4-benzoquinone (system suitability solution), made from USP Reference Standards, at 0.50 mg/L carbon for bulk water

Sucrose and 1,4-benzoquinone, now made from EP CRS (replacing sucrose R and 1,4-benzoquinone R)

0.806 mg/L SDBS solution

System suitability criterion

Response efficiency of 85% to 115%

Same response efficiency calculation as USP, 85% to 115%

The analyser must recover not less than 0.450 mg/L of organic carbon from the SDBS solution

Instrument sensitivity

Limit of detection 0.05 mg/L carbon or lower for bulk water, 0.10 mg/L or lower for sterile water


Able to measure organic carbon down to 0.050 mg/L

Online vs laboratory

Online or offline testing allowed, as long as the result represents the water used

Online or offline testing allowed, as long as the result represents the water used

JP recommends building the instrument into the water production line. Otherwise, test in clean conditions straight after sampling

What changed in EP in 2026

The European Pharmacopoeia Commission adopted revised texts for Water for injections (0169), Water, purified (0008) and general chapter 2.2.44 in June 2025. EDQM confirmed they were published in Issue 12.3 in January 2026 and have applied since 1 July 2026. The changes that matter for TOC are:

  1. New reference substances. Sucrose R and 1,4-benzoquinone R, previously laboratory reagents, were replaced by Chemical Reference Substances (CRS).

  2. TOC for sterilised water for injections. The older oxidisable substances test for sterilised water for injections was replaced by a TOC test, described as Method B in chapter 2.2.44.

  3. Container-dependent limits. Under Method B, the limit depends on the nominal volume of the container: 5 mL or less, over 5 mL to 100 mL, and over 100 mL. Each size has a Limit 1 and a higher Limit 2. A result between the two triggers identification of organic impurities and a risk assessment. A result above Limit 2 fails.

The chemistry and the pass or fail calculation have not changed. What has changed is where the sucrose and 1,4-benzoquinone come from. Veolia explains that USP-sourced materials are now treated as secondary standards for EP purposes, so using them for EP compliance needs a documented equivalence study showing traceability to the EP CRS.

More to come. EDQM also lists a further revised Water for injections (0169) monograph in Issue 13.1, to be implemented on 1 January 2027. Check the official text for what it changes before you update your procedures.

Practical guidance for sites supplying several markets

  1. Map your markets to your pharmacopoeias. List where your products are sold and which pharmacopoeia each regulator expects. Veolia notes the EP requirements apply to anyone selling drug products into European countries, wherever they are made.

  2. Plan for separate USP and EP system suitability sets. If you need both, the simplest route is usually to run USP standards made from USP Reference Standards and EP standards made from EP CRS. If you choose to run only one set, you will need your own documented validation showing equivalence and traceability to the applicable reference standards, and you should expect auditors to ask for it.

  3. Set your test frequency by risk. USP <643> intentionally does not set a frequency for system suitability testing, and Veolia advises setting the EP frequency through a documented risk assessment of your processes and the criticality of the data.

  4. Check packaged water against the right limit. If you fill sterile water into containers, your standards and limits depend on the nominal container size, under both USP and EP.

  5. Check JP separately. If you supply Japan, your system suitability test uses SDBS and your calibration uses KHP. That is a separate set of standards from USP and EP.

  6. Making standards in-house doubles the sourcing. Sites that prepare their own standards can continue to do so, but supplying both USP and EP markets means sourcing sucrose and 1,4-benzoquinone from both USP and EDQM.

  7. Check your analyser with its manufacturer. The EP change is a reference material change, not a technology change, and the instrument manufacturer is the right source for any software or procedure impact.

  8. Keep sampling and handling consistent. Whichever pharmacopoeia applies, clean, low-carbon vials and a consistent sampling routine help keep results meaningful. See our certified TOC vials.

How Sievers analysers and Amscorp TOC standards can support each pharmacopoeia

Sievers TOC analysers. Veolia states that the Sievers M9 TOC Analyzer (Laboratory, Portable and Online) and the Sievers M500 TOC Analyzer need no firmware updates and no changes to calibration or verification procedures to use the new EP standards. Explore our laboratory, online and portable TOC analysers.

USP and EP system suitability standards. Amscorp supplies Sievers system suitability standards in both USP and EP versions, including sets for bulk water and for each sterile water container size. The EP versions are made with EP Chemical Reference Substances. Sites that need both can run the two sets side by side.

JP system suitability standards. Amscorp supplies the Sievers SDBS Suitability Set (STD 90039-01) for customers in Australia and New Zealand. It is designed to meet Japanese and Korean Pharmacopoeia TOC requirements.

Certified and traceable. Amscorp's Sievers standards are supplied with a Certificate of Analysis, and materials for system suitability standards are sourced from the respective pharmacopoeia, traceable to USP, EP, JP and NIST reference materials. See our certified TOC standards.

Frequently asked questions

What is the TOC limit for Purified Water and Water for Injection? USP, EP and JP all work to 0.50 mg/L of carbon for bulk Purified Water and Water for Injection.

Does sterilised water for injections now need TOC testing? Yes, under the European Pharmacopoeia. Since 1 July 2026, the oxidisable substances test for sterilised water for injections has been replaced by a TOC test, with limits that depend on container size. USP has required TOC for sterile packaged waters for some years. EDQM also lists a further revision to the Water for injections monograph (0169) from 1 January 2027, so check the official text when it applies.

Are USP and EP TOC standards still interchangeable? Not directly. Since 1 July 2026, Ph. Eur. 2.2.44 requires system suitability standards made from EP CRS. USP standards used for EP are treated as secondary standards and need a documented equivalence study.

Is the EP pass or fail calculation different from USP? No. Veolia confirms the response efficiency calculation is the same. The difference is the source of the reference materials.

What does JP require for TOC? JP 2.59 calibrates the analyser with potassium hydrogen phthalate and checks it with a 0.806 mg/L sodium dodecylbenzenesulfonate solution, from which the analyser must recover at least 0.450 mg/L of organic carbon.

Can I use one set of standards for USP, EP and JP? Not without your own documented validation. USP, EP and JP each use different reference materials, so most sites supplying all three run separate sets.

How often should I run EP system suitability standards? There is no single answer. Set the frequency through a documented risk assessment of your processes and data criticality.

Does the pharmacopoeia allow online TOC? Yes for USP. USP <643> allows online or offline testing, as long as the result represents the water used. JP 2.59 recommends building the instrument into the water production line. Our laboratory vs online TOC guide explains the trade-offs.

Talk to Amscorp

Working out which standards your site needs for USP, EP or JP? Amscorp Scientific is the exclusive Australia and New Zealand distributor of Sievers TOC analysers and standards, with local AU and NZ support. Book a free product demo or contact our team to talk through your requirements.

Sources

  • United States Pharmacopeia. FAQs: Water for Pharmaceutical and Analytical Purposes (questions on General Chapter <643> Total Organic Carbon). https://www.usp.org/frequently-asked-questions/water-pharmaceutical-and-analytical-purposes. The full chapter text is in the USP-NF (subscription).

  • EDQM. EPC adopts three revised texts related to pharmaceutical waters (18 July 2025). https://www.edqm.eu/en/-/epc-adopts-three-revised-texts-related-to-pharmaceutical-waters-a-major-step-towards-global-quality-standards-for-sterilised-water-for-injections

  • EDQM. Implementation of the European Pharmacopoeia Issue 13.1, notification for CEP holders (23 April 2026). https://www.edqm.eu/en/-/implementation-of-the-european-pharmacopoeia-issue-13.1-notification-for-cep-holders

  • MHLW. The Japanese Pharmacopoeia, 18th Edition, English version: General Tests 2.59 Test for Total Organic Carbon, and Official Monographs for Purified Water, Water for Injection and the packaged waters. https://www.mhlw.go.jp/content/11120000/000945683.pdf and https://www.mhlw.go.jp/content/11120000/000904449.pdf. The same figures appear in the current 19th Edition (published 10 April 2026, Japanese text): https://www.mhlw.go.jp/stf/seisakunitsuite/bunya/0000066530.html

  • Veolia Water Technologies (Sievers). New European Pharmacopoeia EP 2.2.44 TOC Standards: Navigating the Shift and What it Means for Your Facility (28 August 2026). https://www.watertechnologies.com/sievers-resource-center/ep-2-2-44-toc-standards-regulations-update

  • Veolia Water Technologies (Sievers). Application Note 300 40089 EN Rev. A: European Pharmacopoeia (EP) 12.3 Regulatory Update: Adopting New TOC Standards and Container-Dependent Limits. https://estore.watertechnologies.com/document/document/contentdownload/?document_name=ANai_300_40089_EN.pdf&language=English&security=Public

  • Veolia Water Technologies (Sievers). Regulations overview: USP <643> Total Organic Carbon. https://www.watertechnologies.com/sievers-resource-center/usp-643-total-organic-carbon

  • Veolia Water Technologies (Sievers). Changes to USP <643> Total Organic Carbon in Sterile Water (2 April 2025), https://www.watertechnologies.com/sievers-resource-center/usp-643-toc-changes, and Technical Bulletin 300 00262: USP <643> Sterile Water Testing Protocols, https://estore.watertechnologies.com/document/document/contentdownload/?document_name=TBai_300_00262_EN.pdf&language=English&security=Public

  • Veolia Water Technologies (Sievers). Sievers Certified Reference Materials and Consumables Catalog, 300 01338 Rev D (2026). https://estore.watertechnologies.com/document/document/contentdownload/?document_name=Mai_300_01338_EN.pdf&language=English&security=Public

  • Amscorp Scientific. Certified TOC Standards. https://amscorpscientific.com/certified-toc-standards

Precision you can measure.

Service you can count on.

TOC Analysis

Container Closure Integrity Testing

Other Products & Services

Precision you can measure.

Service you can count on.

TOC Analysis

Container Closure Integrity Testing

Other Products & Services

Precision you can measure. Service you can count on.

TOC Analysis

Container Closure Integrity Testing

Other Products & Services